You mean the load to be applied in compression instead tension?If we assume that the load cell works in both direction (tension and compression) and h...

4 days ago
Topic
Reply
RE: Latest software version

In addition, the shunt calibration could be used. It is just a modification (one resistor) of the load cell - that what is used to check the performan...

1 week ago
Reply
RE: RPM sensor issue

Just a quick update... The Opto-Isolator has been ordered and connected - everything works. Thank you!

1 week ago
Reply
RE: Latest software version

@Rub87 for sure, as alternative way, not instead. The reason of direct sensitivity input is to make the calibration easier under certain circumstances...

2 weeks ago
Reply
RE: Latest software version

Hi Jostein, Could it be possible, e.g. in future software releses, to calibrate the Load Cell differently?The load cells, as we use for the testing ...

2 weeks ago
Reply
RE: Display PAD

Please check the version before you buy it - they have Chineese and Inernational. I've read somewhere that Chineese variant doesn't work properly.

2 weeks ago
Topic
2 weeks ago
Replies: 2
Views: 23
Reply
RE: RPM sensor issue

Just tested with with 0.5k - there are 1.938V and 4.2V corresponding.

3 weeks ago
Reply
RE: RPM sensor issue

Hi, I have measured the sensor twice with 1k pullup and without. Without pullup the steady state is 0V, after the optical path triggered the outpu...